Practical Bench / Sensing, Optics & Medical · Beginner · 60 min
Optical fibre loss: cut-back method with plastic fibre
Send red LED light through plastic optical fibre, measure received signal at several lengths, and calculate attenuation in dB per metre.
Why it matters
Internet backbones and hospital endoscopes both depend on how little light is lost per metre. The cut-back method is the standard way to measure it.
Toy or everyday hack
The glowing strands of a toy fibre-optic lamp are real plastic optical fibre
Safety
- Use only the low-power red LED in this practical.
- Never look into any light source through a fibre; this rule matters especially for the invisible infrared lasers used in telecom glass fibre.
- Cut fibre on a mat with a sharp blade away from your fingers, and pick up all offcuts.
- Real telecom glass fibre needs training and laser safety equipment: this practical does not cover it.
Parts list
| Part | Qty | Note |
|---|---|---|
| Plastic optical fibre (1 mm), 5 m | 1 | Or long strands from a toy fibre lamp |
| 5 mm red LED + 330 Ω resistor | 1 | Light source (about 650 nm) |
| Photodiode or phototransistor + 10 kΩ resistor | 1 | Detector |
| 5 V supply or 3 AA batteries (4.5 V) | 1 | |
| Black tube or heat-shrink for light-tight coupling | 1 | About 10 cm each end |
| Sharp blade or fibre cutter, cutting mat | 1 | For clean, flat cuts |
Tools and equipment
Connections
| From | To | Note |
|---|---|---|
| 5 V | 330 Ω → LED → GND | Source |
| 5 V | 10 kΩ → photodiode (reverse biased) → GND | Detector; read voltage across the 10 kΩ |
| LED | Pushed into one end of the fibre tube | Light-tight |
| Fibre far end | Pushed into detector tube | Light-tight |
What you should see
Received light falls steadily as the fibre gets longer, so the dB values fall in a nearly straight line with length. Good 1 mm plastic fibre is specified at roughly 0.15 to 0.25 dB per metre at 650 nm; toy lamp strands and a poor coupling are often much worse. Coupling loss (light that never enters the fibre) is the same for every length, so it shifts the whole line and does not change the slope.
Step by step
- Build the source and the detector. Couple the LED and the photodiode to the ends of the longest fibre with black tubes.
- Cover everything with a cloth to reduce room light. First read the detector voltage with the LED off (dark reading) and enter it in the logbook.
- Switch the LED on and read the detector voltage at full length (for example 5 m). Enter the length and the voltage.
- Cut 1 m off the detector end with a clean flat cut. Re-couple the same way and read again. Repeat until you reach about 0.5 m.
- Do not change anything else (LED current, coupling, cloth) between readings.
- The logbook subtracts the dark reading, calculates the relative power in dB and fits a straight line. The slope is the loss in dB per metre.
- Check how bad a poor cut is: make an angled, rough cut, re-couple and read again.
Challenge: Bend the fibre round cylinders of decreasing diameter and plot the loss against bend radius to see how tight bends leak light.
No code in this practical
This one is done with a multimeter and components only, so you learn the measurement itself. Use the Measure & analyse tab for the calculations.
Readings logbook
Type your measured values. Calculated columns and the analysis update as you type, and your entries are saved in this browser.
If it does not work
| Symptom | Likely cause and fix |
|---|---|
| Readings are noisy | Room light leaking in. Use a darker cover and tape the tubes. |
| Voltage does not follow length | Re-coupling differs every time. Push the fibre in until it stops and keep the same LED distance. |
| Detector saturates (flat top) | Too much light: increase the detector load resistor value or move the LED further away. |
| Curve not straight | Light in the cladding at short lengths or detector near saturation. Use the longer lengths for the slope. |